1. Hydrogen reading higher / lower than expected
Solution:
| |
Ensure alloy calibration constants are set correctly
Ensure analyser is programmed with the correct sensor calibration values |
2. Unusually slow sensor response
Solution:
| |
Clean end of sensor (see manual)
Ensure sensor is stored correctly when not in use. Sensor should not be exposed to ambient humidity for longer than 12h. If sensor will not be used for a long period of time then it should be removed from the probe and stored in a sealed plastic bag containing desiccant. |
3. Erratic / noisy readings
Solution:
4. Unable to enter sensor calibration values
Solution:
| |
Log in as “Engineering”
|
5. Unable to change custom alloy calibration
Solution:
| |
Log in as “Engineering”
|
6. Analyser message: “Temperature too low”
Solution:
| |
The is message is displayed if the probe temperature is < 650 °C
Ensure melt temperature is >650 °C
|
7. Analyser message “Temperature too high”
Solution:
8. Analyser message “Replace Sensor-Failing”
Solution:
| |
Sensor needs replacing
If problem persists then check the probe’s electrical connections (see manual)
|
9. Analyser message “Replace Sensor-Short”
Solution:
| |
Replace sensor (possible aluminium ingress)
If problem persists then check the probe’s electrical connections (see manual)
|
10. Analyser message “Thermocouple failure”
Solution:
| |
Replace thermocouple
|
11. Analyser message “Probe not connected”
Solution:
| |
Connect probe cable
|
12. Unable to activate alarms
Solution:
| |
Alarms are designed for use with a Foundry Degassing Unit and are disabled by default. Contact your Foseco representative for assistance.
|
13. Unable to connect to analyser via PC Bridge program
Solution:
14. Low battery message appears
Solution:
| |
Please contact your Foseco representative to arrange for a new battery to be fitted
|
|